Files found



Current settings


              

Workflow

  1. In the Bio-Rad software (QX ONE or QX600), export the Data Table for each plate as .csv or .xlsx. Raw .ddpcr / .ddpcrone files cannot be read.
  2. Fill in a settings file, one row per well (download the example below).
  3. Upload the exports (a .zip, or the files themselves) and the settings file in the sidebar. Check the files found and the settings check on the Run tab.
  4. Adjust the QC settings if needed, then click Run analysis .
  5. Open the report from the Results tab, or download it with the Excel workbook, RData file and plots.

To try the upload yourself, the demo in the app's demo/ folder has demo_data.zip (the four exports) to upload as the exports and settings/demo_settings.xlsx to upload as the settings file.

Settings file

One row per well. Required columns:

file_name
The plate's export name. Only the stem is used, so run1.ddpcrone , run1.csv and run1.xlsx all match run1 . Several plates can share one settings file.
well
A01 or A1 style.
sample_id
Wells with the same sample_id and target are replicates and are pooled, even across plates. Give samples different IDs if they should not be pooled. IDs are shown exactly as written.
sample_dilution_factor , pcr_sample_vol_ul , pcr_total_vol_ul
Used to correct the concentration back to the original sample: conc × (total vol / sample vol) × dilution factor . Dilution factor 10 for a 1:10 dilution, 1 for none, 0.1 for 10× concentrated.
dye_1 / target_1 , dye_2 / target_2 …
One pair per channel. Dyes are matched to the export's DyeName(s) ignoring case and spaces. A dye with a blank target is treated as an empty channel and left out.
sample_type (optional)
NTC, negative, NRT (no-RT) or blank mark negative controls; anything else (e.g. sample, positive) is analysed as a sample. If the column is missing or blank, the sample type set in the instrument software is used (NTC, NegCtrl).

Well QC

Minimum accepted droplets
Wells with fewer droplets fail. Bio-Rad recommends at least 10,000.
Minimum negative droplets
With no negative droplets the concentration cannot be calculated (Bio-Rad then reports a ceiling value such as 1,000,000 copies/µl).
Maximum positive ratio
Positives / accepted droplets. Default 0.95: with 20,000 droplets the Poisson CV is about 0.9% at 70% positive, 1.0% at 95% and 1.5% at 99%, so precision only collapses as the negatives run out.
Minimum separation
(mean positive amplitude − mean negative amplitude) / mean amplitude of all droplets . Because the mean of all droplets rises with the positive ratio, this index falls as more droplets are positive even when the clusters are equally well apart. The report also gives amplitude_ratio = (positive − negative) / negative, which does not depend on the positive ratio, so the two can be compared across runs before QC is switched to it.
Minimum positives for separation
Separation is only assessed with at least this many positive droplets (default 10). With fewer, the mean positive amplitude rests on a handful of droplets, so the check is left blank and does not fail the well. Wells with zero positives can therefore pass.

Negative controls

A negative control is contaminated for a target when it has more positive droplets than the limit (default 2; one or two stray droplets are common in clean NTCs). Controls are left out of replicate statistics and sample summaries.

Replicate outliers

Each well in a sample × target group of three or more valid replicates is compared with the pooled result of the other wells. It is flagged as an outlier when it differs by more than the fold limit (default 1.5×), its Poisson 95% confidence interval does not overlap the others' (so the gap is more than droplet-counting noise), and the other wells agree with each other. If two of three wells would be flagged, neither is, and the group is marked discordant. Outliers stay in the tables but are left out of the sample means.

Separation and positive-ratio outliers use the Tukey IQR rule within each target, in targets with at least the minimum number of assessed wells.

Outputs

  • kent.html – the report
  • kent.xlsx – all tables; the data sheet is the per-well dataset for downstream work
  • kent.RData – the same tables as R objects ( data_table_processed )
  • plots/ – PNG copies of the main figures

Example settings

Download example settings